94
Keysight N774xC Series Multiport Power Meter Programming Guide
4
Measurement Operations & Settings
Parameters:
DISabled:
AVGover:
MEASure:
Never
When averaging time period finishes
When averaging time period begins
Response:
None
Example:
:trig1:outp dis
Affects:
All power meters
Command:
:TRIGger:CONFiguration?
Syntax:
:TRIGger:CONFiguration?
Description:
Returns the hardware trigger configuration.
Parameters:
None
Parameters:
0 or DISabled:
1 or DEFault:
2 or PASSthrough:
3 or LOOPback:
Trigger connectors are disabled.
The Input Trigger Connector is activated, the incoming trigger response for each slot.
A trigger at the Input Trigger Connector generates a trigger at the Output Trigger
Connector automatically. No triggers from the instrument at the output.
The same as PASSthrough. This is included for compatibility reasons.
Example:
:trig:conf?
->
DEF
Affects
All power meters
Command:
:TRIGger:CONFiguration
Syntax:
:TRIGger:CONFiguration
Description:
Sets the hardware trigger configuration with regard to Output and Input Trigger Connectors.
Parameters:
0 or DISabled:
1 or DEFault:
2 or PASSthrough:
3 or LOOPback:
Trigger connectors are disabled.
The Input Trigger Connector is activated, the incoming trigger response for each
slot.
A trigger at the Input Trigger Connector generates a trigger at the Output Trigger
Connector automatically. No triggers from the instrument at the output.
The same as PASSthrough. This is included for compatibility reasons.
Response:
none
Example:
:trig:conf dis
Affects
All power meters