74
Keysight N774xC Series Multiport Power Meter Programming Guide
4
Measurement Operations & Settings
Response:
Number of loops: Number of loops is an integer value.
0 = Endless Streaming
1 = 1 (Default)
2 = 2 (For 2 Million Points with Buffer A and B) ....
n
Example:
:SENS1:FUNC:LOOP?
→
0
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:PARameter:CHECk?
Syntax:
:SENSe[n]:FUNCtion:PARameter:CHECk?
Description:
Returns whether the currently settings are consistent.
A string with a detailed description of a configuration problem, or "OK" if the sweep os configured correctly. The responses
shown below are all the possible configuration problem string.
Sum of pre trigger and data points is higher than 1048576
No threshold triggering in relative mode allowed
Parameters:
None
Response:
Returns whether the currently settings are consistent.
A string with a detailed description of a configuration problem, or "OK" if the sweep os configured correctly.
The responses shown below are all the possible configuration problem strings:
0,OK
Sum of pre trigger and data points is higher than 1048576
No threshold triggering in relative mode allowed
Example:
:sens1:func:par:check?
→
0,OK
Affects
All power meters
Command:
:SENSe[n]:FUNCtion:PARameter:LOGGing
Syntax:
:SENSe[n]:FUNCtion:PARameter:LOGGing<wsp><data points>, <averaging time>[NS|US|MS|S]
Description:
Sets the number of data points and the averaging time for the logging data acquisition function.