3
Overview
24
N7040A/1A/2A Rogowski AC Current Probes User Guide
Step 1 - Connect to a Compatible Keysight Oscilloscope
Compatibility with Oscilloscopes
To connect to a compatible Keysight oscilloscope
Before you start setting up the probe, ensure that all safety warnings and precautions listed in the
topic
“Pre-Use Safety Checks and Warnings"
on page 20 and 19 are followed.
1 Connect the probe’s BNC output connector to the oscilloscope’s BNC input connector using the
BNC cable provided with the probe.
NOTE
You can use these probes with any Keysight Infiniium or InfiniiVision oscilloscope that has the
1M
Ω
BNC input.
Infiniium Oscilloscopes
InfiniiVision Oscilloscopes
S-Series
6000L Series
9000 Series
6000 X-Series
9000 H-Series
4000 X Series
3000 X Series
3000T X Series
2000 X Series
1000 X Series
1000 Series
Infiniium Software Version
InfiniiVision Software Version
Compatible with all Infiniium
software versions
Recommended version: 6.10 or
higher
Compatible with all InfiniiVision
software versions
WARNING
Grounding is critical to the safe operation of the probe. This connection
using the BNC cable grounds the probe and coil to a properly grounded
oscilloscope.
This connection must always be made BEFORE the coil is wrapped
around the conductor under test and must not be removed until the coil
has been removed from around the conductor under test.
Extreme caution needs to be taken not to damage coil insulation when
placing or removing coil around sharp conductors.