8
User’s and Service Guide N5281-90001
N5281A
General Specifications
General Specifications
Specifications for the N5281A Frequency Converter Test Set (10 MHz to 50 GHz) are
characteristic for the system performance of the PNA and test set. Actual performance of the
system is based on the customers PNA that is used with the test set. A functional certificate is
only offered for the N5281A.
A functional certificate is supplied for the N5281A. The N5281A performance is based on
external components such as the calibration kit, network analyzer, external sources, and
measurement receiver. There are no internal adjustments in the N5281A, therefore an annual
calibration is not required.
Frequency range and connectors are listed in
Table 2
for specific ports.
Table 2
Frequency Range and Connectors
Port
Frequency Range
Connectors
RF Port
0.01 to 50 GHz
2.4 mm female
LO Port
0.01 to 26.5 GHz
3.5 mm female
IF Port
7 MHz
SMA female