Installation Note N5232-90101 19
Step 15.
Reinstall the Outer Cover
For instructions, click the Chapter 7 bookmark “Removing the Covers” in the PDF Service Guide
1
.
Step 16. Enable Option 216
Procedure Requirements
• The analyzer must be powered up and operating to perform this procedure.
• The Network Analyzer program must be running.
• A keyboard must be connected to the network analyzer.
Option Enable Procedure
1. To start the option enable utility, press UTILITY
, then
, then
. An
option enable dialog box will appear.
2. Click the arrow in the
Select Desired Option
box. A list of available options will appear.
3. In the
Select Desired Option
list, click
216 - Configurable Test Set
.
4. Using the keyboard, enter the license key in the box provided. The license key is printed on the the license
message you received from Keysight. Enter this key
exactly
as it is printed on the message.
5. Click
Enable
.
6. Click
Yes
in answer to the displayed question in the
Restart Analyzer?
box.
7. When the installation is complete, click
Exit
.
Option Verification Procedure
Once the analyzer has restarted and the Network Analyzer program is again running:
1. On the analyzer’s
Help
menu, click
About Network Analyzer
.
2. Verify that “216” is listed after “Options:” in the display. Click
OK
.
NOTE
If Option 216 has not been enabled, perform the
again. If the option
is still not enabled, contact Keysight Technologies. Refer to
“Getting Assistance from Keysight”
.
1. See
“Downloading the Online PNA Service Guide” on page 5
System
Service
Option Enable