Keysight N5224B/25B Service Guide 3-47
Tests and Adjustments
Adjustments
Procedure
1.
Connect the equipment as shown in
. Connect a GPIB cable
between the network analyzer and the power meter.
Figure 3-16
Equipment Setup for the Source Adjustment
2.
.Press UTILITY
, then
, then
, then
click Source Adjustment.
3.
Ensure the GPIB settings are correct.
4.
Follow the instructions and prompts as they are displayed.
IF Gain Adjustment
The IF gain adjustment is used to adjust the IF gain of the network analyzer.
Procedure
1.
Press UTILITY
, then
, then
, then
click IF Gain Adjustment.
2.
Follow the instructions and prompts as they are displayed.
Receiver Characterization
This characterizes the receivers in your analyzer.
Procedure
1.
Press UTILITY
, then
, then
, then
click Receiver Characterization.
System