Keysight N5224B/25B Service Guide 4-33
Troubleshooting
Measurement System Troubleshooting
— Connect an Open or Short standard from a mechanical calibration kit to
each test port (use adapters if necessary).
— Press UTILITY
, then ,
then ,
then
.
— For 2-port analyzer models, traces A, B, R1, and R2 are displayed in four
separate data windows as shown in
. Identifying discrepancies of
the traces in these windows can help you to isolate the faulty assembly.
Figure 4-8
Typical 4-Receiver Display for 2-Port Models
— For 4-port analyzer models, traces A, B, C, D, and R1, R2, R3, and R4 are
displayed in eight separate data windows as shown in
.
Identifying discrepancies of the traces in these windows can help you to
isolate the faulty assembly.
System
Service
Utilities
Receiver Display