Service
Module Core Replacement
Ordering a core replacement module
1. Contact Agilent (see
).
2. Order a core replacement for your module (part number M9360-60003).
3. When the core replacement arrives, the package includes:
l
Entitlement Certificate
l
Replacement module
l
Calibration certificate for the replacement module
l
RMA number
l
Return instructions
Replacing the defective module
Before opening a packaged module for troubleshooting, ensure that all ESD (electrostatic discharge)
precautions are observed. Refer to
) for details.
1. Power down the chassis.
2. Remove the defective module from the chassis.
3. Write down the serial number shown on the side shield of the defective module. You will assign this serial
number to the replacement module using the Agilent M9392A Serial Number Update Utility.
4. Remove the replacement module from the box and shipping material.
Agilent M9360A PXI Attenuator/Preselector Service Guide