![Keysight M9202A Startup Manual Download Page 15](http://html1.mh-extra.com/html/keysight/m9202a/m9202a_startup-manual_1961302015.webp)
Step 4: Install the Module
Keysight M9202A PXIe Digitizer, Startup Guide
15
Step 4: Install the Module
PXI hardware does not support "hot-swap" operation. Before
installing the module into the chassis, power off the chassis to
prevent damage to the module.
The module can be installed in any PXIe slot marked with a peripheral slot com-
patibility image (a circle containing the slot number).
The module can also be installed in any hybrid PXIe slot marked with a peripheral
slot compatibility image (the letter “H” and a solid circle containing the slot
number).
1. Make sure the chassis is connected to the main supply, but power switch is in the
Off (Standby) position.
2. If the chassis has multiple fan speed settings, ensure that the fans are set to auto-
matic. Do not set the fan speed to low or turn it off.
3. Position the chassis so that there is ample space between the chassis fan intake
and exhaust vents.
4. If using a remote controller, install the System Interface Card in the slot identified
by the triangle symbol:
5. Install the module into the appropriate slot of the chassis by placing the module
card edges into the front module guides (top and bottom), Ensure that the ejector
handle is pushed down and slide it to the rear.
6. When resistance is felt, latch the module by pulling up on the ejector handle.
Installing a module into a PXI chassis
7. Tighten the screws on the module front panel. Performance may suffer if the
screws are not tightened.