Specifications
M8292A Modular Optical Modulation Analyzer Specifications
Table 4
M8292A Modular Optical Modulation Analyzer specifications
M8292A Modular optical modulation analyzer
Typical values
Maximum detectable symbol rate
74 Gbaud
Sample rate range
83 to 92 GSa/s
Maximum record length per channel
512 kSa
1
ADC resolution
8 bit
Operating frequency range
4
1 MHz to 40 GHz
Analog bandwidth, uncorrected
37 GHz (3 dB)
Optical wavelength operating range
1527.60 to 1570.01 nm
(196.25 to 190.95 THz)
Average input power monitor accuracy
±0.4 dB
Optical phase angle of I-Q mixer after correction
90° ±0.5°
Relative skew after correction
< ±1 ps
Image suppression
5
> 30 dB
Error vector magnitude noise floor
< 2.4% EVM rms at 2.5 GHz freq. offset
5
< 3.5% EVM rms at 10 GHz freq. offset
6
Sensitivity
2
–20 dBm
–14 dBm (with Option 800)
Internal local oscillator (built-in)
Wavelength settling time
< 30 s
Wavelength uncertainty
±4.5 pm (±560 MHz), guaranteed
±2.5 pm (±310 MHz), typical
Sidemode Suppression Ratio (SMSR)
≥
50 dB
Relative Intensity Noise (RIN)
3
–145 dB/Hz (10 MHz to 40 GHz)