34
Keysight M8100A Series Arbitrary Waveform Generators Getting Started Guide
2
Basic Setup for M8100A
Case 3
Connecting M8008A clock module to two M8199A AWG modules
In this case, a M8008A clock module, two M8199A AWG modules, a
M9505A AXIe chassis and the following cables are required:
• Sync cable M8199A-811, (M8199-61620)
• Clock cable (M8199-61624), MWX261 1.85/1.85 (m/m), 450mm
Make the connections as described below:
1 Connect the M8008A Clock Generator Sync Out port with both the
M8199A AWG Sync In ports using the sync cable (blue cable).
2 Connect the M8008A Clock Generator Sample Clock Out port with
both the M8199A AWG Clock In ports using clock cable (red cable).
on page -34 shows how to connect M8008A clock generator
module to the two M8199A AWG modules:
Figure 12
Connecting M8008A clock generator to two M8199A AWG modules
M8008A
M8199A
NOTE
The unused output should be turned off or terminated with 50 Ohms.