17
Keysight M9241/42/43A PXIe Oscilloscopes and M9240A AutoProbe Power Module
Startup Guide
4 Install the Modules
Module Installation Process Overview / 17
Front Panel Features / 19
Cleaning the Instrument / 24
Module Installation Process Overview
The module can be installed in any PXIe or hybrid PXI slot marked with a
peripheral slot compatibility image (solid black circle for PXIe, or solid black circle
with the letter -H- for hybrid).
1
Make sure that the line cord is plugged in to establish earth ground and that
the chassis power switch is
Off
.
2
If the chassis has multiple fan speed settings, ensure that the fan switch is set
to
AUTO
and the inhibit switch is set to
DEF
.
3
Position the chassis to provide ample space between the chassis fan intake and
exhaust vents. Blockage by walls or obstructions affects the air flow needed for
cooling. (Refer to the chassis documentation for more information about
cooling).
4
If using a remote PC controller, install the system interface card into the PC and
connect the chassis to the host PC.
CAUTION
PXI hardware does not support "hot-swap" (changing modules while power is applied
to the chassis) capabilities. Before installing or removing a module to/from the
chassis, power off the chassis to prevent damage to the module.
NOTE
These modules can be used in a chassis with PXIe or PXI-H chassis slots.