Chapter 5: Troubleshooting
To check the LEDs
87
To check the LEDs
If you see a failure with the Auto or Trig'd LEDs, check the voltage at pin 6 of W12,
with W12 disconnected from the keyboard. The voltage should be as follows:
• 0 V
±
0.5 V when both LEDs are supposed to be off.
• 2.5 V
±
0.5 V when Trig'd is supposed to be on and Auto is supposed to be off.
• 5.0 V
±
0.5 V when both LEDs are supposed to be on.
If the voltages are not correct, the problem may be with keyboard cable W12, oscilloscope
interface board A21, acquisition cable W3, or acquisition board A1. Try troubleshooting
the acquisition system first to verify correct behavior before replacing any assemblies.
If the voltages are correct but the LEDs do not light correctly, replace the keyboard
assembly.
If you find a problem with the Armed LED, check pin 5 of W12 with the cable
disconnected from the keyboard. The voltage should be as follows:
• 5.0 V
±
0.5 V when Armed is supposed to be on.
• < 3.6 V
±
0.5 V when Armed is supposed to be on.
Isolation is the same as for the Trig'd and Auto LEDs.
If you find any other failures, replace the keyboard assembly. If the front panel power
indicator LED does not light, replace the cursor keyboard assembly.
Summary of Contents for Infiniium 8000 Series
Page 6: ...6...
Page 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Page 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Page 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Page 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Page 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Page 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Page 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Page 158: ...158...
Page 160: ...160 Index...