![Keysight Infiniium 8000 Series Service Manual Download Page 44](http://html1.mh-extra.com/html/keysight/infiniium-8000-series/infiniium-8000-series_service-manual_1961651044.webp)
Chapter 3: Testing Performance
To test single cursor voltage measurement accuracy with offset
44
Figure 3-3
Voltage Measurement Accuracy Equipment Setup
2
Press Default Setup to set the oscilloscope to default conditions.
3
Set all channels to DC using the Coupling key and to 1 M
Ω
input impedance
using the Input key.
4
Using the mouse, enable the graphical interface.
5
Select Acquisition from the Setup menu.
6
Select Equivalent Time sampling mode. Enable Averaging. Set the # of Averages
to 32, using the numeric keypad dialog. Select Close.
Figure 3-4
Acquisition Setup for Voltage Accuracy Measurement
7
Use the following table for steps 8 through 15.
Select Equivalent Time
sampling mode
Enable averaging
Set the number of
averages to 32
Leave Memory Depth
on Automatic
Summary of Contents for Infiniium 8000 Series
Page 6: ...6...
Page 31: ...Chapter 2 Preparing for Use Digital probe lead set 31 Data Sheets Demonstrations Catalogs...
Page 66: ...Chapter 3 Testing Performance Performance Test Record 66...
Page 140: ...Chapter 7 Replaceable Parts Exploded Views 140 Exploded Views Front Frame and Front Panel...
Page 141: ...Chapter 7 Replaceable Parts Exploded Views 141 Fan and Acquisition Assembly...
Page 143: ...Chapter 7 Replaceable Parts Exploded Views 143 Power Supply and PC Motherboard With Option 017...
Page 144: ...Chapter 7 Replaceable Parts Exploded Views 144 Sleeve and Accessory Pouch...
Page 150: ...Chapter 8 Theory of Operation 150 Figure 8 1 Oscilloscope Block Diagram...
Page 158: ...158...
Page 160: ...160 Index...