128 Keysight E8357D/67D & E8663D PSG User’s Guide
Basic Digital Operation
Triggering Waveforms
—
Segment Ad vance
(Dual ARB only) causes a segment in a sequence to
require a trigger to play. The
trigger source
controls how play moves from
segment to segment (example on
). A trigger received during the
last segment loops play to the
first
segment in the sequence. You have two
choices as to how the segments play:
—
Single
causes a segment in a sequence to play once, then to stop
and wait for a trigger before advancing to the next segment, which
plays to completion. Triggers received during play cause the current
segment to finish, then play advances to the next segment, which
plays to completion.
—
Continuous
causes a segment in a sequence to play continuously
until the waveform receives another trigger. Triggers received during
play cause the current segment to finish, then play advances to the
next segment, which plays continuously.
—
Continuous
repeats the waveform until you turn the signal off or select
another waveform, trigger mode, or response. Continuous has the following
options:
—
Free Run
immediately triggers and plays the waveform; triggers
received during play are ignored.
—
Trigger & Run
plays the waveform when a trigger is received;
subsequent triggers are ignored.
—
Reset & Run
(not available in real–time Custom) plays the waveform
when a trigger is received; subsequent triggers restart the waveform.
Accessing Trigger Utilities
The following figures show the menus for the trigger parameters. These figures
show the Dual ARB player, but you can access the trigger utilities through the
Trigger
softkey in all ARB formats, and the
Pattern Trigger
softkey in the Real
Time I/Q Baseband (Custom) format.
— To display the trigger modes, press
Mode
>
Dual ARB
>
Trigger
.
Active only when
Ext
is selected.
PAT TRIG IN 2 pin on
the AUXILARY I/O connector
PATTERN TRIG IN connector