3 Input/Output Functions
Input/Output
Phase
(Remote Command Only)
This command enables you to select the phase of the V-network that is used, which is controlled via the
AUX IO port. The permissible selection depends on the selected V-network.
Remote Command
INPut[1]|2:LISN:PHASe L1|L2|L3|N
INPut[1]|2:LISN:PHASe?
Example
:INP:LISN:PHAS L1
Couplings
L2, L3 keys are grayed out when ENV216 is selected. If the grayed out key is pressed, an advisory
message is generated. If the equivalent SCPI command is sent, this same message is generated as
part of a “–224, Illegal parameter value; must apply ESH2Z5 to make this phase available” warning.
Preset
Set to N on a "Restore Input/Output Defaults"
State Saved
Saved in instrument state
Range
Phase N|Phase L1|Phase L2|Phase L3
Only one phase can be selected.
Initial S/W Revision
A.14.50
150 kHz Highpass (Remote Command Only)
Controls highpass setting on the V-network.
Remote Command
INPut[1]|2:LISN:FILTer:HPAS[:STATe] ON|OFF
INPut[1]|2:LISN:FILTer:HPAS[:STATe]?
Example
:INP:LISN:FILT:HPAS ON
Dependencies
Only available for ENV216 V-network . This key is grayed out when a V-network that is not ENV216
is selected. If the grayed out key is pressed, an advisory message is generated. If the equivalent SCPI
command is sent, this same message is generated as part of a “–221, Settings conflicts; LISN
function not available” warning.
Preset
Set to off on a "Restore Input/Output Defaults"
State Saved
Saved in instrument state
Range
ON|OFF
Initial S/W Revision
A.14.50
Protective Earth (Remote Command Only)
Enables you to set the Protective Earth setting that is controlled via the AUX IO port.
Remote Command
INPut[1]|2:LISN:PEARth GROunded|FLOating
INPut[1]|2:LISN:PEARth?
Example
:INP:LISN:PEAR GRO
Dependencies
Only available for ESH2Z5. This key is grayed out when a v-network other than ESH2Z5 is selected. If
the grayed out key is pressed, an advisory message is generated. If the equivalent SCPI command is
Sequence Analyzer User's & Programmer's Reference
127