4 Input/Output Functions
Input/Output
Remote Correction Data Set Commands
This section describes the remote (SCPI) commands used to put values into correction sets. See the
correction / table editor section of the Input/Output section for the information on front panel entry of
correction data.
"Set (Replace) Data (Remote Command Only)" on page 215
"Merge Correction Data (Remote Command Only)" on page 215
Set (Replace) Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value separated
by commas.
The values sent in the command will totally replace all existing correction points in the specified set.
An Ampcor array can contain 2000 points maximum.
Remote Command
[:SENSe]:CORRection:CSET[1]|2|...|8:DATA <freq>, <ampl>, . . .
[:SENSe]:CORRection:CSET[1]|2|...|8:DATA?
Example
CORR:CSET1:DATA 10000000, –1.0, 20000000, 1.0
This defines two correction points at (10 MHz, –1.0 dB) and (20 MHz, 1.0 dB) for correction set 1.
Preset
Empty after Restore Input/Output Defaults. Survives a shutdown or restart of analyzer application
(including a power cycle).
State Saved
Saved in instrument state.
Min
Freq: 0 Hz
Amptd: –1000 dBm
Max
Freq: 1 THz
Amptd: +1000 dBm
Initial S/W Revision
A.02.00
Modified at S/W Revision
A.14.00
Merge Correction Data (Remote Command Only)
The command takes an ASCII series of alternating frequency and amplitude points, each value separated
by commas. The difference between this command and Set Data is that this merges new correction points
into an existing set.
Any new point with the same frequency as an existing correction point will replace the existing point’s
amplitude with that of the new point.
An Ampcor array can contain 2000 total points, maximum.
Remote Command
[:SENSe]:CORRection:CSET[1]|2|...|8:DATA:MERGe <freq>, <ampl>, ...
Example
CORR:CSET1:DATA:MERGE 15000000, –5.0, 25000000, 5.0
This adds two correction points at (15 MHz, –5.0 dB) and (25 MHz, 5.0 dB) to whatever values
already exist in correction set 1.
W-CDMA & HSPA Measurement Applications Reference
215
Summary of Contents for E6640A
Page 80: ...Table of Contents lxxx W CDMA HSPA Measurement Applications Reference ...
Page 188: ...3 Programming the Test Set Common Commands 188 W CDMA HSPA Measurement Applications Reference ...
Page 256: ...5 Mode Functions Mode Setup 256 W CDMA HSPA Measurement Applications Reference ...
Page 324: ...6 System Functions System 324 W CDMA HSPA Measurement Applications Reference ...
Page 369: ...8 Channel Power Measurement Auto Couple W CDMA HSPA Measurement Applications Reference 369 ...
Page 389: ...8 Channel Power Measurement Marker W CDMA HSPA Measurement Applications Reference 389 ...
Page 664: ...9 ACP Measurement Auto Couple 664 W CDMA HSPA Measurement Applications Reference ...
Page 669: ...9 ACP Measurement BW W CDMA HSPA Measurement Applications Reference 669 ...
Page 984: ...9 ACP Measurement View Display 984 W CDMA HSPA Measurement Applications Reference ...
Page 1001: ...10 Spectrum Emission Mask Measurement W CDMA HSPA Measurement Applications Reference 1001 ...
Page 1668: ...12 Code Domain Measurement Auto Couple 1668 W CDMA HSPA Measurement Applications Reference ...
Page 2354: ...14 QPSK EVM Measurement 2354 W CDMA HSPA Measurement Applications Reference ...
Page 2364: ...14 QPSK EVM Measurement Auto Couple 2364 W CDMA HSPA Measurement Applications Reference ...