Power supply programming ranges
The
SOURce
subsystem requires parameters for programming
values. The available programming value for a parameter
varies according to the desired output range of the power
supply.
lists the programming values available and
the
MINimum
,
MAXimum
,
DEFault
, and reset values of your
power supply.
Refer to
to identify the programming values when
programming the power supply.
Table 3-2
Keysight E3632A programming ranges
0 to 15 V /7 A range
0 to 30 V /4 A range
Voltage
Programming range
0 V to 15.45 V
0 V to 30.90 V
MAX value
15.45 V
30.90 V
MIN value
0 V
DEFault value
0 V
*RST value
0 V
Current
Programming range
0 A to 7.21 A
0 A to 4.12 A
MAX value
7.21 A
4.12 A
MIN value
0 A
DEFault value
7 A
4 A
*RST value
7 A