Remote Interface Reference
3
Simplified Programming Overview
E3632A User’s Guide
81
Reading a query response
Only the query commands (commands that end with “
?
”)
will instruct the power supply to send a response message.
Queries return either output values or internal instrument
settings. For example, the following statements executed
from your computer will read the power supply’s error
queue and print the most recent error:
Selecting a trigger source
The power supply will accept a bus (software) trigger or an
immediate internal trigger as a trigger source. By default,
the BUS trigger source is selected. If you want the power
supply to use an immediate internal trigger, you must select
IMMediate
. For example, the following statements executed
from your computer will set to an output of 3 V/1 A
immediately:
dimension statement
Dimension string array (80
elements).
SYST:ERR?
Read the error queue.
bus enter statement
Enter the error string into the
computer.
print statement
Print the error string.
VOLT:TRIG 3.0
Set the triggered voltage level to 3.0 V.
CURR:TRIG 1.0
Set the triggered current level to 1.0 A.
TRIG:SOUR IMM
Select the immediate trigger as a source.
INIT
Cause the trigger system to initiate.