268
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
15
Measurements
Occupied Bandwidth
Occupied Band wid th
measures the bandwidth (frequency range) containing some
percent (usually 99%) of the total spectral power. While 99% is the industry norm,
you can specify the percent you want to use in the measurement.
The center frequency used in the measurement is the one defined for the FFT
function, and the FFT span represents the total spectral power.
When this measurement is tracked with cursors, the cursors show the measured
bandwidth (frequency range).
Adjacent Channel Power Ratio (ACPR)
Adjacent Channel Power Ratio or just
Adjacent Power Ratio
(or channel leakage
ratio) measures the ratio of the power contained in one or more sidebands to the
power in the main frequency range.
The main range is specified by a channel width and a center frequency. The
Channel Width is one of the measurement's parameters. The center frequency
used in the measurement is the one defined for the FFT function.
Sidebands (with the same width as the main range) exist above and below the
main range separated by the channel spacing width. The Channel Spacing width
is another measurement parameter.
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...