102
Keysight InfiniiVision 3000T X-Series Oscilloscopes User's Guide
5
Math Waveforms
A scale unit of
U
(undefined) will be displayed for math functions when two source
channels are used and they are set to dissimilar units and the combination of units
cannot be resolved.
Math Operators
Math operators perform arithmetic operations (like add, subtract, or multiply) on
analog input channels.
•
•
Add or Subtract
When you select add or subtract, the
Source 1
and
Source 2
values are added or
subtracted point by point, and the result is displayed.
You can use subtract to make a differential measurement or to compare two
waveforms.
If your waveforms' DC offsets are larger than the dynamic range of the
oscilloscope's input channels you will need to use a differential probe instead.
Math function
Units
add or subtract
V or A
multiply
V
2
, A
2
, or W (Volt-Amp)
d/dt
V/s or A/s (V/second or A/second)
dt
Vs or As (V-seconds or A-seconds)
FFT
See
√
(square root)
V
1/2
, A
1/2
, or W
1/2
(Volt-Amp)
Summary of Contents for DSOX3012T
Page 78: ...78 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 2 Horizontal Controls...
Page 88: ...88 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 3 Vertical Controls...
Page 98: ...98 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 4 FFT Spectral Analysis...
Page 148: ...148 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 7 Digital Channels...
Page 154: ...154 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 8 Serial Decode...
Page 160: ...160 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 9 Display Settings...
Page 232: ...232 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 13 Acquisition Control...
Page 242: ...242 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 14 Cursors...
Page 288: ...288 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 16 Mask Testing...
Page 320: ...320 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 19 Waveform Generator...
Page 340: ...340 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 21 Print Screens...
Page 374: ...374 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 23 Web Interface...
Page 394: ...394 Keysight InfiniiVision 3000T X Series Oscilloscopes User s Guide 24 Reference...