2 Installing the N1000A
Item
Description
Line Power
100/120Vac, 50/60/400 Hz
220/240Vac, 50/60 Hz
Power in Watts
700 Watts Maximum
The products can operate with mains supply voltage fluctuations up to ±10% of the nominal voltage.
Table 3. N1000A LINE Power Specifications
The official and complete specifications are located in the N1000A’s help
system. In FlexDCA, click Help > Specifications….
Use Keysight supplied power cord or one with same or better electrical rating.
To avoid injury, use a cart or similar device when transporting the instrument.
When lifting the instrument, use safe lifting techniques. Lift with your legs, not
with your back. Always employ at least two persons who have been trained in
safe-lifting techniques.
The Mains wiring and connectors shall be compatible with the connector used
in the premise electrical system. Failure, to ensure adequate earth grounding
by not using the correct components may cause product damage, and serious
injury.
Install the instrument so that the detachable power cord is readily identifiable
and is easily reached by the operator. The detachable power cord is the
instrument disconnecting device. It disconnects the mains circuits from the
mains supply before other parts of the instrument. The front panel switch is only
a standby switch and is not a LINE switch. Alternatively, an externally installed
switch or circuit breaker (which is readily identifiable and is easily reached by
the operator) may be used as a disconnecting device.
Windows registry: If the N1000A is mounted in a rack or cabinet, do not use the
system power switch to disconnect power from the instrument. Instead, use the
N1000A front-panel power switch. Using the system power switch may corrupt
the Windows registry requiring you to perform the recovery procedure to
restore normal operation to the N1000A.
Should the Declaration of Conformity (DOC) be required, contact an Keysight
Sales Representative or the closest Keysight Sales Office. Alternately, contact
www.keysight.com.
24
N1000A Quick Start Guide
Summary of Contents for DCA-X Series
Page 1: ...Keysight N1000A DCA X Series Wide Bandwidth Oscilloscope Quick Start Guide...
Page 7: ...1 Introduction Front Panel Features Figure 3 Front Panel Features N1000A Quick Start Guide 7...
Page 8: ...1 Introduction Rear Panel Features Figure 4 Rear Panel Features 8 N1000A Quick Start Guide...
Page 39: ...3 Removable Hard Disk 39 N1000A Quick Start Guide...
Page 47: ...5 Contacting Keysight 47 N1000A Quick Start Guide...