Quick-Start
1
Getting Started Guide
25
AC power cord
The test set is equipped with a three-wire power cord, in accordance with international
safety standards. This cable grounds the test set cabinet when connected to an
appropriate power line outlet. The cable appropriate to the original shipping location is
included with the test set. See:
http://www.agilent.com/find/powercords
Environmental Conditions (Operating)
This product is designed for use in the following conditions:
•
For indoor use only
•
Temperature 5°C to 50°C
•
Altitude up to 3048 meters
•
0% Relative Humidity to 90% Relative Humidity up to 40°C, and linearly decreasing
to 50% Relative Humidity up to 50°C; non-condensing..
Always use the three-prong AC power cord supplied with this product.
Failure to ensure adequate earth grounding by not using this cord can
cause product damage.
This instrument has auto-ranging line voltage input; be sure the supply
voltage is within the specified range and voltage fluctuations do not to
exceed 10 percent of the nominal supply voltage:
100/120 VAC 50/60 Hz,
220/240 VAC 50/60 Hz,
270 W Max.
If this product is not used as specified, the protection provided by the
equipment could be impaired. This product must be used in a normal
condition (in which all means for protection are intact) only.
Install the instrument so that the detachable power cord is readily
identifiable and easily reached by the operator. The detachable power
cord is the instrument disconnecting device. It disconnects the mains
circuits from the mains supply before other parts of the instrument.
The front panel switch is only a standby switch and is not a LINE
switch. Alternatively, an externally installed switch or circuit breaker
(which is readily identifiable and is easily reached by the operator)
may be used as a disconnecting device.
This product is designed for use in Installation Category II and
Pollution Degree 2.