ACQuire:CCDF:FRAMe <numeric_value>
This command sets the CCDF frame.
Syntax
Parameter
Example
NOTE
The CCDF frame can only be set when the CCDF triggered sweep mode is
enabled (refer to
“ACQuire:CCDF:SWEep <character_data>”
ACQ
:CCDF
Space
?
:FRAM
numeric_value
Item
Description/Default
Range of values
numeric_value
A numeric value for the CCDF frame:
– Default value: 1
– Minimum value: 1
– Maximum value: 1G
The range of values above is determined by CCDF trigger duration
×
CCDF
frame and must be
≤
100 s.
1 to 1 G
ACQ:CCDF:FRAM 2K
This command sets the CCDF frame to 2K.