OCS{1|4}?
This query returns the Operating Calibration Status of channel 1 or channel 4 on
the 8990B.
– 1 (true) is returned if channel 1 or 4 is calibrating
– 0 (false) is returned if channel 1 or 4 is not calibrating
Syntax
Example
Remark
– This query is only applicable for channels 1 and 4. If a channel other than
channel 1 or 4 is specified, the following errors — “–113,“Undefined
header;OCS2<Err>” and “–420,”Query UNTERMINATED” — will occur.
– This query is only applicable when calibration is invoked via SCPI commands.
OCS1
?
1
4
OCS1?
Queries the Operating Calibration Status of channel 1 on the
8990B.