![Keysight 8710 Series Operating And Service Manual Download Page 24](http://html1.mh-extra.com/html/keysight/8710-series/8710-series_operating-and-service-manual_1961606024.webp)
3
Specifications
24
Keysight 8710x Operating and Service Manual
RF Specifications
Table 3-2
RF specifications for 87104P/Q and 87106P/Q
Keysight model number
87104P and 87106P
87104Q and 87106Q
Frequency range
DC to 4 GHz
DC to 20 GHz
Insertion loss (dB)
0.015f
[a]
+ 0.3
[a] f = frequency in GHz
0.015f
+ 0.3
Isolation (dB)
> 100
> 100 (DC to 12 GHz)
> 80 (12 to 15 GHz)
> 70 (15 to 20 GHz)
SWR
< 1.20
< 1.20 (DC to 4 GHz)
< 1.35 (4 to 12.4 GHz)
< 1.45 (12.4 to 20 GHz)
Repeatability
[b]
(dB)
[b] Up to 3 million cycles, measured at 25 °C
< 0.03
< 0.03
Characteristics
50
Ω
, terminated
50
Ω
, terminated
Connectors
SMA (f)
SMA (f)
Table 3-3
RF specifications for 87104R and 87106R
Keysight model number
87104R and 87106R
Frequency range
DC to 26.5 GHz
Insertion loss (dB)
0.015f
[a]
+ 0.3
[a] f = frequency in GHz
Isolation (dB)
> 100 (DC to 12 GHz)
> 80 (12 to 15 GHz)
> 70 (15 to 20 GHz)
> 65 (20 to 26.5 GHz)
SWR
< 1.20 (DC to 4 GHz)
< 1.35 (4 to 12.4 GHz)
< 1.45 (12.4 to 20 GHz)
< 1.70 (20 to 26.5 GHz)
Repeatability
[b]
(dB)
[b] Up to 3 million cycles, measured at 25 °C
< 0.03
Characteristics
50
Ω
, terminated
Connectors
SMA (f)
Summary of Contents for 8710 Series
Page 1: ...Keysight 8710x Low PIM Coxial Multiport Switches Operating and Service Manual ...
Page 4: ...4 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 6: ...6 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 8: ...8 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 10: ...10 Keysight 8710x Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...