Preparing for Imaging
5
Keysight 5500 SPM User’s Guide
5-2
voltages to opposite piezo elements in the scanner so that one element
elongates and the other contracts.
The scanner mounting fixture supplied with your system is designed to
keep the scanner and its components safe during handling (
The main cutout safely holds the scanner, while the smaller cutout
safely holds a nose assembly. A magnetic disk keeps additional tools
close at hand.
Figure 4-1
Scanner mounting fixture with nose assembly and spring
key; scanner in mounting fixture
The next sections will describe how to safely handle the scanner
components for long life and excellent imaging.
One-Piece Nose Assembly
Inserting the One-Piece Nose Assembly
The nose assembly is held in the scanner by an O-ring around its
circumference. To insert a nose assembly, first place the scanner in the
scanner mounting fixture (
). Place the nose assembly in the
socket on top of the scanner, aligning its contact pins if applicable.
CAUTION
The thickness of the piezo elements determines how much they will
expand or contract per applied unit voltage. They are necessarily thin to
provide scanning resolution.
If dropped, the piezo elements inside the
scanner WILL break.
Cracked or broken piezoelectrodes will result in
abnormal imaging. Proper handling is essential to preserve the long
expected life of your multi-purpose scanner.
CAUTION
Ensure that the ends of the metal spring for retaining the cantilever are
not caught in between the nose cone and the socket.