Scanner Maintenance and Calibration
5
Keysight 5500 SPM User’s Guide
5-14
If the measured dimension does not match the actual, then adjust the X
Sensitivity term according to the following equation:
After adjustment the measured size should be within 5 % of the actual
size.
Y Calibration
The methods used for calibrating the scanner’s Y axis are similar to
those used for the X dimension. The exception is that the scan range will
be set to ½ of the scanner’s full range. The Y axis is generally used as
the slow scanning axis so the range will be reduced as a time
consideration.
Only one Topography image is required for Y calibration. The other
image can be assigned to display flattened Topography data.
Y Non-Linearity
Obtain an image of the calibration target. Using the Cross Section tool,
set markers at the uppermost and bottommost feature along a vertical
cross section (
Figure 8-11
Cross-section in Y direction showing non-linearity
NewSensitivity
CurrentSensitivit y Kno
wnSize
MeasuredSize
--------------------------------------------------------------------------------------
=