Keysight 42030A Four-Terminal Pair Standard Resistor Set
13
Calibration and Repair
Calibration
5 MHz
± 0.20%
± 0.20
Ω
10 MHz
± 0.5%
± 0.5
Ω
13 MHz
± 0.7%
± 0.7
Ω
Table 2-5
42037A 1 k
Ω
Resistor AC Calibration Uncertainty
Frequency
Calibration Uncertainty
Parallel Resistance
Susceptance
100 kHz
± 0.12%
± 0.8 µS
1 MHz
± 0.08%
± 1.0 µS
2 MHz
± 0.10%
± 1.0 µS
3 MHz
± 0.10%
± 1.5 µS
4 MHz
± 0.15%
± 2.0 µS
5 MHz
± 0.15%
± 3.0 µS
10 MHz
± 0.4%
± 5 µS
13 MHz
± 0.6%
± 8 µS
Table 2-6
42038A 10 k
Ω
Resistor AC Calibration Uncertainty
Frequency
Calibration Uncertainty
Parallel Resistance
Susceptance
100 kHz
± 0.08%
± 0.10 µS
1 MHz
± 0.08%
± 0.20 µS
Table 2-7
42039A 100 k
Ω
Resistor AC Calibration Uncertainty
Frequency
Calibration Uncertainty
Parallel Resistance
Susceptance
100 kHz
± 0.08%
± 0.02 µS
1 MHz
± 0.12%
± 0.15 µS
Table 2-4
42036A 100
Ω
Resistor AC Calibration Uncertainty (Continued)
Frequency
Calibration Uncertainty
Series Resistance
Reactance