66
4000 X-Series Oscilloscopes Advanced Training Guide
3
Advanced Triggering, Search & Navigation, and Segmented Acquisition Labs
27
Press the “runt polarity” softkey and select the “either polarity” (third) icon.
28
Tap the
Qualifier
softkey to select
None
.
29
Press the
[Acquire]
front panel key.
30
Tap the
Segmented
softkey; then double-tap
# of Segs
to enter
500
as the
number of segments to capture.
31
Tap the
Segmented
softkey to begin a Segmented Memory acquisition.
32
Tap the
Current Segment
softkey; then turn the Entry knob to review all 500
captured segments.
33
Set the
Current Segment
to
500
and note the time-tag of the last captured
segment.
Segmented Memory optimizes oscilloscope acquisition memory by only capturing
important segments of a waveform based on the trigger condition and timebase
setting. In this example, we have selectively captured 500 occurrences of positive
and negative runts of any width for a total acquisition time of more than 20 ms as
shown in
Figure 40
Using the scope’s Segmented Memory acquisition to selectively capture 500
consecutive runt pulses.