Introduction to the Plug-In Modules for the 34980A
4
Keysight 34980A Mainframe User’s Guide
219
Electrical Operating Conditions
Transients
The 34921A, 34922A, 34923A, 34924A, 34925A, 34931A, 34932A, 34933A,
34934A, 34937A, 34938A and 34939A modules are designed to safely withstand
occasional transient overvoltages up to 1000 Vpeak. Typically, these transient
overvoltages result from switching inductive loads or from nearby lightning
strikes. The lightning-caused transient overvoltages that may occasionally occur
on mains power outlets may be as high as 2500 Vpeak.
34941A
Four channels, 30V, 0.5A,
10W per channel
Four channels, 30V, 0.5A,
10W per channel
34942A
Four channels, 30V, 0.5A,
10W per channel
Four channels, 30V, 0.5A,
10W per channel
34945A
See 34945A User’s Guide
See 34945A User’s Guide
34946A
Dual channel, 7V, 1W per channel,
4 GHz or 20 GHz
Dual channel, 7V, 1W per channel,
4 GHz or 20 GHz
34947A
Triple channel, 7V, 1W per channel,
4 GHz or 20 GHz
Triple channel, 7V, 1W per channel,
4 GHz or 20 GHz
34950A
64 channels, 5V, 30 mA Max
64 channels, 5V, 30 mA Max
34951A
4 channels, 16V, 20 mA
4 channels, 16V, 20 mA
34952A
32 DIO channels, 42V, 400 mA,
2 channel DAC, 12V, 10 mA
32 DIO channels, 42V, 400 mA,
2 channel DAC, 12V, 10 mA
34959A
See 34959A User’s Guide
See 34959A User’s Guide
Environmental Operating Limits (current and power dissipation) for the Plug-In
Modules (continued)
Module
Pollution Degree 1 Specifications
Pollution Degree 2 Specifications
WARNING
To avoid electric shock, turn off the 34980A and disconnect or de-energize
all field wiring to the modules and the Analog Bus connector before
removing any module or slot cover.