General Information
1
Keysight 11683A Operating and Service Manual
19
Description
The 11683A Range Calibrator is used to verify proper operation of compatible Power Meters such
as the Keysight Model 435A. The Power Meter's range
to-range accuracy and proper auto-zero
operation can be easily verified. The 11683A can supply a full-scale test signal to the Power Meter
for each Range Switch setting.
When set to CALIBRATE, the FUNCTION switch applies a dc voltage to the Power Meter; the input
is grounded in STANDBY. The POLARITY switch increases ease of testing and adjusting the Power
Meter auto-zero feedback circuit.
Figure 1-2
Four-wire cable
Table 1-3
Recommended test equipment
Instrument
Critical specification
Model
Use
Digital voltmeter
Readout: 5 digits
DC measurement
Range: 100 mV to 100 V full-scale
Accuracy: ± 0.02%
Resistance measurement (four-wire measurement capability)
Range: 100
Ω
to 10 k
Ω
full-scale
Sensitivity: 1mil
Accuracy: ± 0.02%
3455A
Performance,
Adjustment,
Troubleshooting
Oscilloscope
Vertical amplifier
Bandwidth: DC to 5 MHz
Deflection factor: 50 mV /division minimum
Attenuator accuracy:± 2%
Time Base
– Time span/division: 1 ms to 1 s
– Time base accuracy: ± 3%
180C/1801A/1821A
Adjustment,
Troubleshooting
Four-wire cable
Recommended length: 5 feet maximum
See
Performance
Summary of Contents for 11683A
Page 1: ...Keysight 11683A Range Calibrator Operating and Service Manual...
Page 12: ...12 Keysight 11683A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 14: ...14 Keysight 11683A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK...
Page 51: ...Service 6 Keysight 11683A Operating and Service Manual 51 Figure 6 11 11683A FG top view...