Y1277A Installaton Guide                                                                                                                                    27

Safety and Maintenance Information
Instrument Maintenance

Instrument Maintenance

Cleaning the instrument

To remove dirt or dust from the external case of the M9484C VXG, clean the 

case using a dry or slightly-dampened cloth only.

Cleaning Connectors

Cleaning connectors with alcohol shall only be done with the instrument power 

cord removed, and in a well-ventilated area. Allow all residual alcohol moisture 

to evaporate, and the fumes to dissipate prior to energizing the instrument.

Keep isopropyl alcohol away from heat, sparks, and flame. Store in a 

tightly closed container. It is extremely flammable. In case of fire, use 

alcohol foam, dry chemical, or carbon dioxide; water may be ineffective.
Use isopropyl alcohol with adequate ventilation and avoid contact with 

eyes, skin, and clothing. It causes skin irritation, may cause eye damage, 

and is harmful if swallowed or inhaled. It may be harmful if absorbed 

through the skin. Wash thoroughly after handling.
In case of spill, soak up with sand or earth. Flush spill area with water.
Dispose of isopropyl alcohol in accordance with all applicable federal, 

state, and local environmental regulations.

Summary of Contents for Y1277A

Page 1: ...Installation Guide Keysight Y1277A Kit Removable SSD for the M9484C...

Page 2: ...o U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license...

Page 3: ...tps www keysight com us en product M9484C m9484c html Information on preventing instrument damage can be found at http keysight com find PreventingInstrumentRepair Is your product software up to date...

Page 4: ...4...

Page 5: ...Setting the Instrument s Computer Name 17 Installing License Files 20 3 Safety and Maintenance Information Safety Information 22 Warnings Cautions and Notes 23 General Safety Considerations for the M9...

Page 6: ...6 Y1277A Installaton Guide...

Page 7: ...lowing topics can be found in this section Overview on page 8 Putting the Additional SSD Into Service on page 9 SSD Interconnect Life Expectancy on page 10 SSD Removal and Installation on page 11 Firm...

Page 8: ...r Y1277A along with your M9484C VXG follow the instructions below The SSD is mounted in a receptacle on the rear panel of the M9484C chassis and can be easily inserted or removed This is convenient wh...

Page 9: ...stalled in the instrument The instruments computer name Licenses associated to this particular instrument When ordered after receipt of the instrument the following information must be obtained from t...

Page 10: ...fe expectancy of both the connector on the removable SSD and the mating connector internal to the instrument is 100 cycles If this is exceeded it is very likely that a failure will occur between these...

Page 11: ...at the SSD and remove the SSD 4 Install the new SSD by aligning the SSD PC board edges with the card guides Insert the SSD completely 5 Tighten the retained screw to no more than 9 in lbs Electrostati...

Page 12: ...led instrument firmware version on the VXG select the Triple bar icon Settings Firmware Version If an instrument firmware update is required the latest revision of the firmware can be found by going t...

Page 13: ...13 Installing the Y1277A SSD Instrument Security Information Instrument Security Information Information on the instrument security features and the instrument volatility can be found at http www key...

Page 14: ...14 Y1277A Installaton Guide Installing the Y1277A SSD Instrument Security Information...

Page 15: ...for Use in the M9484C The Keysight Y1277A provides an additional removable Solid State Drive SSD for your M9484C VXG The following procedures will accomplish the initialization process Setting the In...

Page 16: ...laton Guide Initializing the Y1277A SSD Accessory for Use in the M9484C Requirements Requirements PathWave signal generator version A 09 00 or higher running on the SSD Mouse and keyboard connected to...

Page 17: ...ows Start menu and select the Change Account settings icon and sign out 2 At the Login prompt enter the user name and password The factory default settings are User name administrator Password Keysigh...

Page 18: ...the System Properties dialog select Change 7 In the Computer Name Domain Changes dialog enter your desired Computer Name The default factory Computer name is K M9484C 12345 where 12345 is the last 5 d...

Page 19: ...in the M9484C Setting the Instrument s Computer Name 8 Select OK You will now be prompted to restart the instrument Make sure to save any open files and close all programs then select OK to continue...

Page 20: ...Files Copy the license file s from the original SSD onto the additional SSD Retrieve the files from where you have stored them off instrument and copy then to this folder on the instrument C ProgramD...

Page 21: ...tenance Information The following topics can be found in this section Safety Information on page 22 Warnings Cautions and Notes on page 23 Instrument Markings on page 26 Instrument Maintenance on page...

Page 22: ...ividual instruments to avoid having the heated discharge of one instrument now becoming the cooling intake air for another instrument Another area of concern is verification that the maximum ambient o...

Page 23: ...tion to a condition or situation that could result in personal injury or loss of life Do not proceed beyond a warning until you fully understand the indicated conditions or situations Caution calls at...

Page 24: ...switch or circuit breaker which is readily identifiable and is easily reached by the operator may be used as a disconnecting device This is a Safety Protection Class I Product provided with a protect...

Page 25: ...rounding by not using the correct components may cause product damage and serious injury The main power cord can be used as the system disconnecting device It disconnects the mains circuits from the m...

Page 26: ...Class A product CISPR 11 Clause 5 Korean Certification KC mark includes the markings identifier code which follows the format R R Kst ZZZZZZZ This WEEE symbol indicates separate collection for electr...

Page 27: ...to energizing the instrument Keep isopropyl alcohol away from heat sparks and flame Store in a tightly closed container It is extremely flammable In case of fire use alcohol foam dry chemical or carb...

Page 28: ...www keysight com find assist Americas Canada 1 877 894 4414 Latin America 305 269 7500 United States 1 800 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India...

Page 29: ...ired Call your Keysight Technologies office for full details If you want to service the instrument yourself after the warranty expires you can download the service documentation that provides all nece...

Page 30: ...This information is subject to change without notice Keysight Technologies 2022 Edition 1 March 2022 Y1277 90001 www keysight com...

Reviews: