Keysight CXG, EXG, and MXG X-Series Signal Generators Programming Guide 69
Programming Examples
GPIB Programming Interface Examples
Remote Function
The HP Basic function
REMOTE
and the other listed IO library functions change
the signal generator from local operation to remote operation. In remote
operation, the front panel keys are disabled except for the
Local
key and the
line power switch. Pressing the
Local
key restores manual operation.
VISA Library
In VISA, the viTerminate command requests a VISA session to
terminate normal execution of an asynchronous operation. The
parameter list describes the session and job id.
viTerminate
(parameter list)
NI-488.2
The NI-488.2 library function aborts any asynchronous read,
write, or command operation that is in progress. The
parameter
ud
is the interface or device descriptor.
ibstop(int ud)
SICL
The Keysight SICL function aborts any command currently
executing with the session
id
. This function is supported with
C/C++ on Windows 3.1 and Series 700 HP-UX.
iabort (id)
Library
Function Statement
Initialization Command
Library
Function Statement
Initialization Command
HP Basic
The
REMOTE 719
function disables the front panel
operation of all keys with the exception of the
Local
key.
10 REMOTE 719
VISA Library
The VISA library, at this time, does not have a similar
command.
N/A
NI-488.2
The NI-488.2 library function asserts the Remote Enable
(REN) GPIB line. All devices listed in the parameter list are put
into a listen-active state although no indication is generated
by the signal generator. The parameter list describes the
interface or device descriptor.
EnableRemote
(parameter list)
SICL
The Keysight SICL function puts an instrument, identified by
the
id
parameter, into remote mode and disables the front
panel keys. Pressing the
Local
key on the signal generator
front panel restores manual operation. The parameter id is the
session identifier.
iremote (id)
Summary of Contents for X-Series