Data Questionable Calibration Event Enable Register
The Data Questionable Calibration Event Enable Register lets you choose
which bits in the Data Questionable Calibration Event Register set the
summary bit (bit 8 of the Data Questionable Condition register) to 1.
Command:
STATus:QUEStionable:CALibration:ENABle <value>
, where
<value>
is the sum of
the decimal values of the bits you want to enable.
Example:
Enable bit 1 and bit 0 so that whenever either of those bits are set to 1, the Data Questionable
Calibration summary bit of the Data Questionable Condition Register is set to 1. Send the command
STAT:QUES:CAL:ENAB 3
(2 + 1).
Query:
STATus:QUEStionable:CALibration:ENABle?
Response:
Decimal value of the sum of the bits previously enabled with the
STATus:QUEStionable:CALibration:ENABle <value>
command.