182 Keysight CXG, EXG, and MXG X-Series Signal Generators Programming Guide
Programming the Status Register System
Status Groups
Standard Event Status Group
The Standard Event Status Group is used to determine the specific event that
set bit 5 in the Status Byte Register. This group consists of the Standard Event
Status Register (an event register) and the Standard Event Status Enable
Standard Event Status Register
NOTE:
Some
signal
generator
models may
use only a
subset of
the status
registers
Bit
Description
0
Operation Complete
. A 1 in this bit position indicates that all pending signal generator operations were completed
following execution of the
*OPC
command.
1
Request Control
. This bit is always set to 0. (The signal generator does not request control.)
2
Query Error
. A 1 in this bit position indicates that a query error has occurred. Query errors have instrument error
numbers from
−
499 to
−
400.
3
Device Dependent Error
. A 1 in this bit position indicates that a device dependent error has occurred. Device
dependent errors have instrument error numbers from
−
399 to
−
300 and 1 to 32767.