Making Connections
3
Keysight V281C/D and W281C/D Operating and Service Manual
29
Cleaning Connectors
Supplies and equipment needed
Contact Keysight Technologies (
on page 3) for a list of the supplies
and equipment that are needed to perform the cleaning procedures, and their Keysight
Technologies part numbers.
Basic cleaning procedure
1
Inspect the connectors for dust, dirt, metal fragments, oils or films, and debris.
2
Blow off any dust with a filtered, clean supply of compressed air.
3
Add a few drops of high-purity isopropyl alcohol to small cleaning swab (do not apply alcohol
directly to the parts).
4
Gently
wipe connecting surfaces with the end of the cleaning swab (see
).
5
Blow dry with a clean supply of compressed air.
6
Inspect and repeat cleaning procedure if necessary.
Figure 3-7
Cleaning illustration
NOTE
A minimum magnification of 10X is required for all inspections of the adapter parts.
NOTE
When using isopropyl alcohol to clean connectors
do not allow
liquid to flow down inside the
connector. This may cause measurement errors due to residue inside the connector. If possible
keep the connector facing down.