U3047AM08
Cal Kit Operational Check
Figure
40
Calibration Complete
6.
Repeat steps
thru
for Ports 2 thru 12. When finished, there should be 16
Cal Sets saved with the titles “999.1” thru “999.12” (12-Port).
7.
If you are using an ECal module you can verify the individual port calibration by
selecting Response > Cal > More > ECal > ECal Confidence Check - OR-
(Response > Cal > Cal Sets & Cal Kits > ECal Confidence Check)
. Select
Change Measurement and select the test port
S-Parameter > Apply > OK >
Read Module Data
. For more information, refer to the Help menu.
8.
Select
Trace/Chan > Trace > Delete Trace
. There should be no traces on the
network analyzer screen.
9.
To launch the Cal Set Viewer toolbar. Select
Response > Cal > Manage Cals >
Cal Set Viewer
.
Figure
41
Calibration, Cal Set Viewer
10.
From the Cal Sets drop-down menu, select
999.1
and select
Enable
. Select
the
Reflection Tracking(x,x)
, where x,x is the port being tested, term in the
center
drop-down menu and ensure that the Enable and Error Terms check
boxes are
selected.
11.
Compare the Reflection Tracking (1,1) trace to the appropriate limits in
. The trace should be above the limit.
12.
Repeat steps
10 and
1
1
for Cal Sets “999.1” thru “999.12” (12-Port).
Keysight U3047AM08 User's and Service Guide
59