U3047AM08
Cal Kit Operational Check
-
1-Port Calibration and Verification Procedure
1.
Connect the ECal or the mechanical cal kit to Port 1 or the port you are
testing. Torque to 8 in-lb. For further information refer to the Help menu,
.
2.
Perform a 1-Port Calibration on Port 1. On the network analyzer, select
Response > Cal > Start Cal > Calibration Wizard -OR-
(Response > Cal > Other Cals)
.
a.
If using a mechanical cal kit, select
SmartCal (Guided Calibration) >
Next.
b.
If using an ECal module, connect the ECal to a network analyzer
USB port. Select
Use Electronic Calibration (ECal) > Next
.
3.
Continue following the Cal Wizard prompts. In the Select Calibration Ports
and ECal Module dialog box click
Clear All
, then select
Port 1 > Next >
Measure
.
Figure 3
9
1-Port Calibration
4.
Continue to follow the prompts until the Calibration Completed dialog box
appears.
5.
At the Calibration Completed prompt, select
Save As User Calset
, type the
name
999.1
. Overwrite the Calset if it already exists and
Save
.
Help
If you do not have a keyboard, select
Save As User Calset > Edit Name
and save as
999.x
. X is the port number you are calibrating. Use the
numeric keypad on the network analyzer front panel to enter “999.1.”