86
Keysight U3042AM20 User’s and Service Guide
U3042AM20
Troubleshooting the Test Set
-
Noise Source Out Test
Ext Noise Source Out to Source In (near Port 5)
1. Set the PNA to Standalone Mode.
2. Connect two RF test cables to the PNA-X Ports 1 and 2. Connect the two
cables together using an adapter.
3. Perform a response calibration to normalize and remove the loss created
by the test cables.
4. Remove the adapter.
5. Connect the cable from Port 1 to Ext Noise Source Out.
6. Connect the cable from Port 2 to the Test Set Source In.
7. Send the Test Set I/O command 0.8. Verify that the Ext Noise Source front
panel LED
is On.
8. Set trace S21 and verify that the trace is similar to
.
Figure 58
Noise Source Out to Source In Path Response
Summary of Contents for U3042AM20
Page 7: ...Keysight U3042AM20 User s and Service Guide 1 U3042AM20 ...
Page 105: ......