Network Analyzer Requirements
Ensure that the network analyzer has the latest version of installed. The following
web site links provide the necessary information needed:
• Network Analyzer -
http://www.keysight.com/find/pna
• U3042AE12 Test Set Files -
http://na.support.keysight.com/multiport
• Firmware -
http://na.support.keysight.com/p
Table
2
All network analyzers require Option 551 for multiport operation (N-Port error
correction and measurement capability).
provides a listing of other network
analyzer option requirements. The analyzer’s frequency range must be equal to or
exceed that of the test set so that the measurement system covers the full frequency
range of the test set.
The test set files indicated in
must be installed into the analyzer’s file
directory: c:\Program Files\Keysight\Network Analyzer\testsets
Network Analyzer Configuration Requirements
4-Port
Network Analyzer
Options
Test Set File
System Figure
N5222A/B PNA
401, 417 or 419
u3042ae12_pnax_p4.tsx
N5230A/C PNA-L
245, 145, 246 or 146
u3042ae12_p4.tsx
N5232A/B PNA-L
416
u3042ae12_p4.tsx
N5242A/B PNA-X
400
u304xae12_pnax_p4.tsx
10
U3042AE12 User's and Service Guide