System Configurations
The Keysight U3022BH10 High Power Test Set can be configured for many applications.
Included in this document are four typical configurations:
•
System Setup, 14 Ports" on page 2
•
System Setup, 14/24 Ports with U3020AH37" on page 31
•
, “1 Watt High Power Configuration”
•
, “20 Watt High Power Configuration”
Refer to section,
“High Power Measurements” on page
The internal firmware of the
Keysight
N5242A
/B
PNA-X Series Network Analyzer
has not been modified for this Test Set option. Power levels may differ from
those indicated on the PNA-X when the Test Set is connected.
Refer to the configuration diagrams for external component connections and/or
operating constraints when utilizing the high power capability. External
components are not supplied with the high power Test Set. See
and
The high power Test Set is equipped with receiver access ports. The jumpers can
be removed to insert attenuators that will reduce the RF power to the PNA-X A,B
and R channel receivers. Recommended power levels to the PNA-X receiver
ports is –15 dBm. Refer to your PNA-X specifications to optimize power levels to
the receiver ports.
Prior to powering-up the booster amplifier it is highly recommended that the
user verify the RF power levels seen by the various elements of the Test Setup.
Both the PNA-X and Test Set power levels must be taken into consideration. At
high power levels, a mistake could damage the instruments. Accordingly, the
following key values are given in
22