![Keysight Technologies U3022AM06 User'S And Service Manual Download Page 64](http://html1.mh-extra.com/html/keysight-technologies/u3022am06/u3022am06_users-and-service-manual_1962380064.webp)
Control Lines
U3022AM06
Control Lines
The 15 pin female D-Sub connector on the rear panel provides 8 latched data lines
that can be used to control your device under test (DUT). The lines can be controlled
with the Multiport External Test Set control, or Test Set I/O commands. Refer to
“External Test Set Control Feature” on page
“Setting the Control Lines with
Address and Data Values” on page 6
Table 16
for DUT control line specification.
The output voltage of the lines can be from the internal adjustable voltage source
(+2 to +5 Vdc), or an external DC power supply depending on how the connection to
the control line is configured. When using an external power supply a positive or
negative voltage can be used. Refer to
for control line pin location and description. Refer to
66
and
“External Voltage Supply Configuration” on page
for configurations.
Table 16
DUT Control Specifications
Item
Specifications
Connector Type
15–pin female D–Sub
Max Output Line Current
100 mA (each line)
Control Line DC resistance
< 10
Ω
(each line)
Voltage Range:
Positive Input
0 to +5 V
Negative Input
–5 to 0 V
Internal Variable Voltage
+2 to +5 V
64
Keysight U3022AM06 User's and Service Guide