safety precautions required to avoid
possible injury. Read and follow all
installation, operation, and
maintenance information carefully
before using the product.

If this product is not used as specified,
the protection provided by the
equipment could be impaired. This
product must be used in a normal
condition (in which all means for
protection are intact) only.

The types of product users are:

Responsible body is the individual or
group responsible for the use and main-
tenance of equipment, for ensuring that
the equipment is operated within its spe-
cifications and operating limits, and for
ensuring operators are adequately trained.

Operators use the product for its intended
function. They must be trained in electrical
safety procedures and proper use of the
instrument. They must be protected from
electric shock and contact with hazardous
live circuits.

Maintenance personnel perform routine
procedures on the product to keep it oper-
ating properly (for example, setting the line
voltage or replacing consumable mater-
ials). Maintenance procedures are
described in the user documentation. The
procedures explicitly state if the operator
may perform them. Otherwise, they should
be performed only by service personnel.

Service personnel are trained to work on
live circuits, perform safe installations, and
repair products. Only properly trained ser-
vice personnel may perform installation
and service procedures.

Operator is responsible to maintain safe
operating conditions. To ensure safe
operating conditions, modules should
not be operated beyond the full
temperature range specified in the
Environmental and physical
specification. Exceeding safe operating
conditions can result in shorter
lifespans, improper module
performance and user safety issues.
When the modules are in use and
operation within the specified full

temperature range is not maintained,
module surface temperatures may
exceed safe handling conditions which
can cause discomfort or burns if
touched. In the event of a module
exceeding the full temperature range,
always allow the module to cool before
touching or removing modules from
chassis.

Keysight products are designed for use
with electrical signals that are rated
Measurement Category I and
Measurement Category II, as described
in the International Electrotechnical
Commission (IEC) Standard IEC 60664.
Most measurement, control, and data
I/O signals are Measurement Category I
and must not be directly connected to
mains voltage or to voltage sources with
high transient over-voltages.
Measurement Category II connections
require protection for high transient
over-voltages often associated with
local AC mains connections. Assume all
measurement, control, and data I/O
connections are for connection to
Category I sources unless otherwise
marked or described in the user
documentation.

Exercise extreme caution when a shock
hazard is present. Lethal voltage may
be present on cable connector jacks or
test fixtures. The American National
Standards Institute (ANSI) states that a
shock hazard exists when voltage levels
greater than 30V RMS, 42.4V peak, or
60VDC are present. A good safety
practice is to expect that hazardous
voltage is present in any unknown
circuit before measuring.

Operators of this product must be
protected from electric shock at all
times. The responsible body must
ensure that operators are prevented
access and/or insulated from every
connection point. In some cases,
connections must be exposed to
potential human contact. Product
operators in these circumstances must
be trained to protect themselves from
the risk of electric shock. If the circuit is
capable of operating at or above 1000V,
no conductive part of the circuit may be
exposed.

Do not connect switching cards directly
to unlimited power circuits. They are

intended to be used with impedance-
limited sources. NEVER connect
switching cards directly to AC mains.
When connecting sources to switching
cards, install protective devices to limit
fault current and voltage to the card.

Before operating an instrument, ensure
that the line cord is connected to a
properly-grounded power receptacle.
Inspect the connecting cables, test
leads, and jumpers for possible wear,
cracks, or breaks before each use.

When installing equipment where
access to the main power cord is
restricted, such as rack mounting, a
separate main input power disconnect
device must be provided in close
proximity to the equipment and within
easy reach of the operator.

For maximum safety, do not touch the
product, test cables, or any other
instruments while power is applied to
the circuit under test. ALWAYS remove
power from the entire test system and
discharge any capacitors before:
connecting or disconnecting cables or
jumpers, installing or removing
switching cards, or making internal
changes, such as installing or removing
jumpers.

Do not touch any object that could
provide a current path to the common
side of the circuit under test or power
line (earth) ground. Always make
measurements with dry hands while
standing on a dry, insulated surface
capable of withstanding the voltage
being measured.

The instrument and accessories must
be used in accordance with its
specifications and operating
instructions, or the safety of the
equipment may be impaired.

Do not exceed the maximum signal
levels of the instruments and
accessories, as defined in the
specifications and operating
information, and as shown on the
instrument or test fixture panels, or
switching card.

When fuses are used in a product,
replace with the same type and rating
for continued protection against fire
hazard.

3

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for U1084A

Page 1: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Page 2: ... bandwidth Notice This document contains references to Agilent Please note that Agilent s Test and Measurement business has become Keysight Technologies For more information go to www keysight com Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 3: ...Accordingly Keysight provides the Software to U S government customers under its standard commercial license which is embodied in its End User License Agreement EULA a copy of which can be found at http www keysight com find sweula The license set forth in the EULA represents the exclusive authority by which the U S government may use modify distribute or disclose the Software The EULA and the lic...

Page 4: ...urces unless otherwise marked or described in the user documentation Exercise extreme caution when a shock hazard is present Lethal voltage may be present on cable connector jacks or test fixtures The American National Standards Institute ANSI states that a shock hazard exists when voltage levels greater than 30V RMS 42 4V peak or 60VDC are present A good safety practice is to expect that hazardou...

Page 5: ...로 합니 다 This product complies with the WEEE Directive marketing requirement The affixed product label above indicates that you must not discard this electrical electronic product in domestic household waste Product Category With reference to the equipment types in the WEEE directive Annex 1 this product is classified as Monitoring and Control instrumentation product Do not dispose in domestic house...

Page 6: ...re Windows 12 Installation Procedure Linux 13 Step 4 Install the Module 13 Reversing the orientation of the fan assembly 14 U1084A Front Panel Features 15 Front Panel Connectors 15 STATUS LED 15 Step 5 Using the Applications 16 Running the MD1 SFP Application 16 Windows 16 Linux 16 Running C C LabVIEW or MATLAB Applications under Windows 17 New C C software applications 17 LabVIEW applications 18 ...

Page 7: ...Documentation Map Documentation Map 6 Keysight U1084A Startup Guide Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 8: ...ur local Keysight representative or contact us through our website at www keysight com find contactus Related Documentation This Startup Guide and the documentation listed below are contained on the CD supplied with your product and at www keysight com find U1084A Select Document Library Manuals U1084A User Manual Help system for the Soft Front Panel Help systems for the Keysight device drivers IV...

Page 9: ...nal verification of the U1084A PCIe High Speed Digitizer Step 1 Unpack and Inspect Step 2 Verify Shipment Step 3 Install Drivers and Software Step 4 Install Modules Closely follow the startup process flow in this document Deviating from the sequence can cause unpredictable system behavior damage your system and may cause personal injury 8 Keysight U1084A Startup Guide Artisan Technology Group Qual...

Page 10: ...n Both types when used together provide a significant level of ESD protection Of the two only the table mat and wrist strap combination provides adequate ESD protection when used alone To ensure user safety the static safe accessories must provide at least 1 MΩ of isolation from ground DO NOT use these techniques for a static safe work station when working on circuitry with a voltage potential gre...

Page 11: ...top cover of the module The serial number can also be read from the Soft Front Panel interface but only after the hardware is installed Description of failure or service required 4 Pack the module in its original ESD bag and packing carton If the original carton is not available use bubble wrap or packing peanuts and place the instrument in a sealed container and mark the container FRAGILE 5 On th...

Page 12: ...Super VGA graphics is supported Does not require graphics headless system X Windows with 1280x1024 recommended for SFP Browser Microsoft Internet Explorer 7 0 or higher Distribution supplied browser 1 Because of the installation procedure less disk space may be required for operation than is required for installation The amount of space listed above is required for installation 2 NET Framework Run...

Page 13: ... digitizer Keysight IO Libraries Suite is not required 4 After installation is complete please shut down the PC The MD1 Software Package provides a specific fundamental driver AgMD1Fundamental dll which will be installed in the C Program Files IVI Foundation IVI Bin folder In addition it also provides the driver contained in the previous Acqiris Software Package AqDrv4 dll allowing backward compat...

Page 14: ...that the PC is switched off and disconnect the power cord Open the PC case 2 Follow ESD precautions when handling and installing the U1084A 3 When installing the U1084A ensure that it has optimum cooling do not install it in a position where the integrated cooling fans may be obstructed 4 Loosen but do not remove the 4 screws holding the rear support bracket shown in violet in the diagram and slid...

Page 15: ... be positioned side by side To change the orientation please use the following procedure 1 Remove the 2 screws top bottom nearest to the front of the fan unit and remove the 2 retaining clips shown pale red in the diagram 2 Slide the fan unit approximately 15mm to the rear unit it clears the circuit board and rotate it by 180 3 Slide the fan unit back towards the module until it is fully seated in...

Page 16: ...V DC The input full scale ranges are selectable from 50 mV to 5V Bandwidth depends on the option 001 DC to 1 5 GHz 1 8 GHz typ 002 DC to 1 5 GHz 1 8 GHz typ 003 DC to 500 MHz typ TR OUT MCX female A trigger signal output DC coupled Amplitude 0 8 V Output impedance 50 Ω I O A B C MCX female User configurable multi purpose input out put signals 3 3 V CMOS and TTL com patible STATUS LED LED State Des...

Page 17: ...the introduction of the MD1 SFP 5 x However at this point in time the SFP 5 application does not fully support all products modes of operation or features that are supported by the SFP 1 x and so the user must decide which SFP to use depending on their application MD1 SFP 5 x The Soft Front Panel SFP 5 can be launched from Start All Programs Agilent MD1 MD1 SFP or MD1 SFP_64 MD1 SFP 1 x This appli...

Page 18: ... your project to use the new AgMD1Fundamental driver as follows 1 In the project properties set the include directory to C Program Files x86 Agilent MD1 U10xx_Digitizers include default installation 2 In the project properties link to the new library AgMD1Fundamental lib in place of AqDrv4 lib or AgMD1Fundamental_64 lib in place of AqDrv4_64 lib These libraries are located in C Program Files x86 A...

Page 19: ... tar LinuxMD1 X xx xx U10xx Examples New C C software applications Please refer to the programming information contained in the README file section 6 2 An example program and makefile is located in usr share doc md1 examples Existing C C software applications If you have previously written applications with the Acqiris software it will be necessary to make some changes to your project to use the n...

Page 20: ...Keysight U1084A Startup Guide 19 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 21: ...ubject to change without notice Keysight Technologies 2011 2015 Edition 2 December 2015 Printed In USA U1084 90002 www keysight com Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 22: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Reviews: