Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data”. Enter the data
from the external DMM. Press Enter when done. Press
Back to finish.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Current Temperature Coefficient Calibration
The temperature coefficient calibration procedure must be performed Before any
other current calibration procedures.
Step 1. Connect a precision shunt resistor to the output. The shunt resistor should be able to measure
the output's full-scale current (see
). Connect the Keysight 3458A DMM across the shunt
resistor.
Step 2. Select the temperature coefficient calibration.
Front Panel Menu Reference
SCPI Command
Select
System\Admin\Cal\Misc\CurrTC
.
Check that the shunt is connected and select Next.
Step 3.Select the first current calibration point. Wait 5 minutes for the temperature to settle.
Calculate the shunt current (I=V/R) and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P1 measured data”. This should be
about 50% of the full-scale current rating. Press Enter when
done.
CAL:LEV P1
*OPC?
CAL:DATA <data>
Step 4. Select the second current calibration point. Wait 5 minutes for the temperature to settle.
Calculate the shunt current (I=V/R) and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data”. Enter the data from
the external DMM. This should be about 80% of the full-scale
current rating. Press Enter when done.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Step 5. Select the third current calibration point.Wait 5 minutes for the temperature to
settle.Calculate the shunt current (I=V/R) and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P3 measured data”. Enter the data
from the external DMM. This should be about 100% of the
full-scale current rating. Press Enter when done. Press
Back to finish.
CAL:LEV P3
*OPC?
CAL:DATA <data>