CALibrate Subsystem
Calibrate commands calibrate the instrument.
Read the
before calibrating. Improper calibration reduces
accuracy and reliability.
CALibrate:COUNt?
Returns the number of times the unit has been calibrated. The count is incremented whenever the
calibration (and date) is saved, the administration password is changed or reset, or the firmware is
updated.
Parameter
Typical Return
(none)
<count>
Return the calibration count: CAL:COUN?
CALibrate:CURRent[:LEVel] <
value
>
Calibrates the current programming and measurement. The value selects the range to calibrate.
Parameter
Typical Return
The maximum current of the output range.
(none)
Calibrates the current of the 10 A range: CAL:CURR 10
CALibrate:CURRent:SHARing
Calibrates the Imon signal for paralleled units.
Parameter
Typical Return
(none)
(none)
Calibrates the current sharing: CAL:CURR:SHAR
CALibrate:CURRent:TC
Calibrates the temperature coefficient.
Parameter
Typical Return
(none)
(none)
Calibrates the temperature coefficient: CAL:CURR:TC
Keysight RP7900 Series Operating and Service Guide
179
5 SCPI Programming Reference