17
Getting Started
Power On and Check
Check for Instrument Messages
The spectrum analyzer has two categories of instrument messages:
error and warning messages. A error message is triggered by
operation errors, for example, parameter setting conflicts or data
input that is out of the range of a parameter. An warning message
may be triggered by hardware defects which could result in
damage to instrument.
Here are some tips to check the instrument messages.
1
Check the display to see if any messages display in the status bar.
Press
[System] > System Info > Error History
to review each
message. Refer to
Chapter 5, "Instrument Messages"
for detailed
system message descriptions.
2
When you have reviewed and resolved all of the error messages,
press
[System] > System Info > Error History > Clear error queue
to
delete the messages.
3
Cycle the power to the analyzer and re-check to see if the
instrument messages are still there.
4
If the error messages cannot be resolved, please contact the
Keysight customer contact center for assistance or service.