16
Installation Note N9038-90039
Option EXM – External Mixing Upgrade
20.
Referring to
, orient W46 so that the cable slopes upward from the J3 connector at a
15 degree angle. The top of the connector should be approximately level with the top of the
shield over the front panel interface board. Torque the connection to J3 to 10 inch-lbs. using a
5/16" open end wrench.
Figure 9
W46 Cable Orientation