N9030B PXA Signal Analyzer Service Guide 247
Analog/Digital IF Troubleshooting
A3 Digital IF Troubleshooting
10.
The analyzer marker should read 100 MHz at
+
10 dBm
±
4 dB as shown in
.
Figure 7-12
100 MHz Reference Input
11.
If the 100 MHz signal is measuring incorrectly, see
Assembly Troubleshooting” on page 268.
in this service guide.
12.
If the 22.5 MHz IF, 250 MHz IF (if appropriate) and 100 MHz reference
signals measure the correct frequency and amplitude and yet the display
is not processing the signal properly, the most probable causes are the A3
Digital IF or the A4 CPU. It is difficult to separate these two assemblies
given the architecture of the analyzer. Each assembly will have to be tried
to see which one is causing the failure.