194 N9030B PXA Signal Analyzer Service Guide
RF Section Troubleshooting (Millimeter-Wave Analyzers)
Troubleshooting
To perform the YTF alignment, reconnect W62 or W57 cable, and press
System, Alignments, More, Ad vanced, Characterize Preselector
. The routine
may take several minutes to align the YTF. Display the 4.8 GHz calibrator
signal on screen as explained in the quick check section. If the signal level is
still incorrect, suspect all assemblies, cables, switches and the A15 Reference
calibrator signal between the W62/W57 cable and the input to the A9
Attenuator.
If Option MPB is not present, remove A10 Input Attenuator B output cable and
measure the output power. Expected signal is 4.8 GHz
−
38 dBm. See
for location of attenuator. If the attenuator output signal is correct suspect
the A11 Low Band Switch, and if Option MPB or LNP is installed, suspect
switches 3 and 4. Also suspect the switching signals from the A15 Front End
controller. The control signals are explained in the Front End Control
Troubleshooting section.
Figure 5-8
RF Section
A9 Input Attenuator A and A10 Input Attenuator B Verification
Calibrator Switch Test
On the A16 Reference assembly, disconnect semi rigid cable W3 from A16J701
and measure A16J701 with a spectrum analyzer. Expected signal is 4.8 GHz at
−
28 dBm
±
0.5 dB. If signal level is incorrect, suspect A16 Reference Assembly.
Summary of Contents for N9030B
Page 4: ...4 ...
Page 14: ...14 N9030B PXA Signal Analyzer Service Guide Contents ...
Page 342: ...342 N9030B PXA Signal Analyzer Service Guide Hardware Options Option RTS Streaming ...
Page 344: ...344 N9030B PXA Signal Analyzer Service Guide Block Diagrams Block Diagrams Block Diagrams ...
Page 572: ...572 N9030B PXA Signal Analyzer Service Guide Post Repair Procedures Post Repair Procedures ...
Page 612: ...612 N9030B PXA Signal Analyzer Service Guide Instrument Software Software Updates ...